TomoJAX alignment estimates geometry and per-view pose corrections while it reconstructs the volume. It is experimental: on every scan, check the recovered parameters as well as the image.

Start from corrected line integrals and checked physical geometry, following the real scan guide or the lab cone-beam CT guide. Save an unaligned reconstruction to compare with. Commands run from the checkout root.

Choose an alignment mode

ProblemModeCommand
Choose a data-dependent workflow and inspect its limitationsauto (the default)tomojax align scan.nxs -o aligned.nxs
The sample moved from projection to projectionposetomojax align scan.nxs -o aligned.nxs --mode pose
The detector centre (centre of rotation) is wrongcortomojax align scan.nxs -o aligned.nxs --mode cor
A detector-centre offset and per-view motion togethercor-then-posetomojax align scan.nxs -o aligned.nxs --mode cor-then-pose
Mild setup error (centre, detector roll, axis direction) and motionfulltomojax align scan.nxs -o aligned.nxs --mode full
A multi-orbit cone-beam scan, or a scan that already carries posesposetomojax align scan.nxs -o aligned.nxs --mode pose
The reference elevation or a detector-v shift is uncertainNone: check it by handdet_v_px is not a reliably recoverable alignment target

On cone-beam data, cor estimates the rotation axis's offset and the detector roll instead of a detector shift; see cone-beam scans.

Run an alignment

uv run --no-sync tomojax align corrected.nxs -o aligned.nxs --mode pose
uv run --no-sync tomojax inspect aligned.nxs
uv run --no-sync tomojax recon aligned.nxs -o recon.nxs --method cgls

Automatic alignment records its selected workflow and any limitations in result.info["automatic"]; the CLI prints limitations without requiring a manifest. These conditional checks do not certify physical pose truth.

The aligned file holds the corrected scan, its per-view poses and the volume the alignment converged with, and tomojax recon applies the poses. tomojax align corrects on top of any poses saved in its input; with either command, --no-poses starts from the nominal geometry instead.

tomojax align is tj.align on the command line: its options are tj.align's keywords (--mode, --quality, --levels, --freeze, --grid, --checkpoint). --quality reference runs slower, more conservative solver settings, and --levels 4 2 1 sets the coarse-to-fine binning factors. --dry-run prints the resolved plan for an explicit mode as JSON without aligning. Automatic mode reports that calibration and region selection are data-dependent; it does not run them during a dry run.

Expert settings are tomojax.alignment.AlignConfig fields: put them in a TOML file and pass it with --config FILE and an explicit --mode. Each replaces that field of the configuration the mode and quality give, which --dry-run prints; tomojax align --config-keys lists the fields with the pose profile's values. The alignment solver reference describes them.

In Python:

from dataclasses import replace

import tomojax as tj
from tomojax.alignment import alignment_plan

scan = tj.load("corrected.nxs")
result = tj.align(scan, mode="pose")
tj.save("aligned.nxs", result)
recon = tj.reconstruct(result.scan, "cgls")  # result.scan carries the corrections

# An expert setting: replace one field of the mode's configuration.
config = replace(alignment_plan("pose", scan.grid).config, ray_integrator="exact")
result = tj.align(scan, mode="pose", config=config)

scan.with_poses(table) attaches externally estimated or interpolated corrections in Scan.poses's detector frame; it replaces, rather than adds to, existing poses. scan.with_poses(None) removes poses without undoing setup calibration.

result.poses is the (views, 6) pose table and result.info the solver's record.

Complex ptychography projections

For an axial parallel scan of complex transmission, unwrap its phase and remove an air-supported phase plane before treating the negative phase as line integrals. The ptychography example shows bounded preparation, pixel-unit geometry, ProjectionMatchingConfig and final Fourier reconstruction in physical units when the input supplies a wavelength and pixel size. The smaller simulated PSI MATLAB v5 example has no such scale; its CPU recipe reports phase per original pixel instead. See the example commands.

This translation-only configuration jointly estimates a sparse volume and per-view motion, then attaches the poses without resampling measured projections. Individual vertical shifts can be wrong by tens of pixels even when the image and half-dataset FSC improve. On already aligned input, fitting noise as motion can worsen the reconstruction. Compare with the unaligned image and inspect individual pose estimates; FSC and held-out residuals are diagnostics, not proof of physical accuracy. The stopping tolerance is not a pose-error bound.

With ProjectionMatchingConfig, per-view vertical standard deviations are in result.info["vertical_uncertainty"]["standard_deviation"], in the detector's length units. They describe the local vertical fit with its profile template and noise covariance held fixed. They do not cover a wrong initial alignment or image-model bias. Horizontal and full-procedure uncertainties are not reported.

result.info["initialization"]["losses"] contains the final losses from the two starts. Similar losses can indicate an ambiguous horizontal fit; they are not pose-error bounds.

Align large scans at reduced resolution

Pose parameters are physical lengths and angles, so they transfer between resolutions. For a large scan, stop the coarse-to-fine schedule early and reconstruct the full data with the recovered poses:

uv run --no-sync tomojax align corrected.nxs -o aligned.nxs \
  --mode pose --levels 4 2
uv run --no-sync tomojax recon aligned.nxs -o recon.nxs \
  --method fista --nonnegative --tv-weight 0 --iterations 300

Laminography leaves a cone of frequencies unmeasured, so the full-resolution reconstruction needs a prior: positivity, as here, and TV (--tv-weight) for noisy data. On a 256³ laminography scan, stopping at half resolution took a quarter of the time for slightly less accurate rotations (measurements). When the detector samples finer than the grid, binning it also helps; see large scans.

Align on a region or another grid

grid= aligns and reconstructs on another grid than the scan's, a region around the object or coarser voxels, as it does for tj.reconstruct. The returned volume is on that grid and result.scan carries it. Use an explicit mode with this override: automatic laminography selects its own grid and region.

region = tj.Grid(192, 192, 128, scan.grid.vx, scan.grid.vy, scan.grid.vz)
result = tj.align(scan, mode="pose", grid=region)
assert result.scan.grid == region

On the command line, --roi (auto, off, cube, bbox, cyl) selects the field of view; --grid NX NY NZ resamples that physical extent to the requested voxel counts. In Python, scan.grid.resampled((NX, NY, NZ)) makes a coarser or finer grid of the same extent to pass as grid=.

Resume an interrupted alignment

With checkpoint=, alignment saves its progress (volume, poses, solver state) after each outer iteration. Run the same call again after an interruption and it resumes where the last checkpoint left off, reaching the result an uninterrupted run would; once the run has finished, the call returns its result at once:

result = tj.align(scan, mode="pose", checkpoint="align.ckpt")

A checkpoint resumes only the alignment that wrote it: the same projections and geometry (checked by a fingerprint of sixteen evenly spaced views and of every geometry setting, angles and pose corrections included), mode, levels, grid and AlignConfig, written by the same TomoJAX version. Anything else raises ValueError naming the difference, for example config differs in freeze (checkpoint [], current ['dx']), and the file is left as it is: delete it or choose another path to start again. Checkpoints written by versions before schema 4 do not resume.

tomojax align --checkpoint PATH does the same: run the same command again to resume.

Use several GPUs

tj.align(scan, mode="pose", devices=jax.devices()) shares the views of the joint update and of cone-beam reconstruction steps among the GPUs; the rest runs on the first. A checkpoint written on some GPUs resumes on others. The command line uses one GPU. See large scans and several GPUs.

Check the result

Alignment can improve the image without every recovered parameter being a physical calibration: pose-only correction can absorb some setup error, so the reconstruction can look good while the poses differ from the true geometry. Neither a lower objective nor a sharper image proves that the estimated geometry is calibrated.

  • Compare the aligned reconstruction with the unaligned one in several slice planes, and check the residuals.
  • Check recovered setup parameters against what you know of the acquisition. To estimate a detector-centre offset explicitly, use cor or cor-then-pose rather than reading it from pose-only results.
  • In pose mode the offset stays in the per-view dx; tj.align reports the implied offset, in detector pixels, as result.info["implied_detector_u_px"], and tomojax align logs it and writes it to the --manifest.

Known failure modes:

  • Abrupt jumps may need jump-aware pose handling, and short bursts of bad views a robust loss or bad-view detection.
  • A per-view shift with a nonzero mean over the scan cannot be told apart from a detector-centre offset; cor-then-pose assigns it to the offset.
  • Detector-v or sample-elevation reference shifts are physically ambiguous and are not reliably recoverable.
  • Shifts beyond a quarter of the detector, which leave the object partly outside the field of view, are not recovered.
  • Large combined setup and pose errors can need staged initialisation, stronger priors or manual review.

Correct per-view motion

Explicit pose mode optimises one pose per projection: rotations alpha, beta and phi, and translations dx and dz. Pose tables carry a sixth column, dy, a translation along the beam. It has no effect on parallel-beam projections and stays at zero there; cone-beam alignment estimates it. Five- column tables from earlier versions load with dy = 0.

Each Gauss–Newton step updates the volume and the poses together, using Joseph plane sampling and an unregularised least-squares fit. Up to 30 outer iterations stop early once the fit stops improving. Scans of 64³ and larger align coarse to fine by default. Accuracy depends on resolution: coarse grids leave a discretisation floor on the rotations, even when the solve starts from the true poses (measurements).

Large per-view stage shifts are found first by a global shift search (the seed_translations setting, on by default in pose mode): reconstruct with the current shifts removed, reproject, and move each view to its cross-correlation peak, searching up to a quarter of the detector. Local solvers alone converge only from shifts of a few pixels.

The setting gn_coupling = "fixed_volume" restores the older scheme that refines poses against a reconstruction held fixed between volume updates. It accepts other losses, smooth pose models and optimisers, but on the free-voxel pilot it left rotation errors of 0.1–1° that the coupled solver removes.

Smooth pose models

The default pose model is per_view, which optimises an independent pose for every projection. Use a smooth model when you expect the motion to change smoothly over the scan. Smooth models need the alternating solver; put both in a TOML file, here spline.toml:

gn_coupling = "fixed_volume"
pose_model = "spline"
knot_spacing = 8
uv run --no-sync tomojax align corrected.nxs -o aligned.nxs --mode pose --config spline.toml

Smooth models reduce degrees of freedom but can hide abrupt jumps or outlier views.

Calibrate the detector centre

For an unposed axial parallel scan covering a half-turn, tj.estimate_center gives a projection-only estimate with no volume solve:

scan = tj.estimate_center(tj.load_frames("raw.nxs").corrected())
recon = tj.reconstruct(scan)

It returns a new Scan whose detector.center[0] is the absolute physical detector-u coordinate, estimated from up to three detector rows. The Fourier metric can have false minima for compact objects, so check the reconstruction.

Use cor mode when the main problem is a detector-u or centre-of-rotation offset rather than sample motion:

uv run --no-sync tomojax align corrected.nxs -o aligned.nxs --mode cor

COR mode fits the detector-u offset explicitly. It starts from a one-parameter search for the offset whose FBP reprojects most consistently, which works for laminography and partial arcs, then refines it against held-out views. The search assumes the views are otherwise consistent: under per-view motion it misjudges the offset, so use cor-then-pose when the sample also moves. The measurements give its accuracy.

To compare trial centres by eye instead, see the quickstart.

Estimate a detector offset together with motion

cor-then-pose runs the same solver as pose and then reports the constant part of the recovered detector-u shifts as the detector centre. With detector-frame translations (the default) a detector-u offset is exactly a constant dx; a rigid object translation instead shifts each view by a sinusoid of its angle, so a fit over the scan separates the two. The output geometry carries the offset as Detector.center, and the saved dx values are the remaining per-view motion; together they predict the same data as the alignment, so the volume is unchanged. full adds the same constant to the detector centre from its setup stage.

uv run --no-sync tomojax align corrected.nxs -o aligned.nxs --mode cor-then-pose

In the Python API, align_multires with AlignConfig(schedule="cor_then_pose", pose_translation_frame="detector") returns the offset as det_u_px in info["geometry_calibration_state"], and implied_detector_offset performs the separation on any pose table.

Use mixed setup and pose as expert mode

full mode combines setup and pose stages: detector centre, detector roll and axis direction, then per-view motion, coarse to fine. Because setup and pose parameters can represent similar image changes, mixed correction has gauge ambiguity, and each stage handles it with a fixed gauge policy: the pose stage anchors the mean translation (anchor_mean) and the axis-direction stage is diagnostic only (diagnose_only). --dry-run prints each stage's policy.

uv run --no-sync tomojax align corrected.nxs -o aligned.nxs --mode full

An expert direct parameter set that mixes setup and pose parameters (the optimise_dofs setting) must choose a policy itself, with the gauge_policy setting, in a file passed with --config:

optimise_dofs = ["det_u_px", "alpha", "beta", "phi", "dx", "dz"]
gauge_policy = "anchor_mean"

Gauge policies:

  • anchor_mean: anchors the mean translation so setup and pose do not drift together; this is a convention, not an independent calibration.
  • prior_required: requires physical setup priors.
  • diagnose_only: produces diagnostics without treating the result as a calibrated correction.
  • reject: fails instead of running an ambiguous mixed correction.

Align cone-beam scans in six degrees of freedom

In a cone beam, moving the sample along the beam changes its magnification, so tomojax align on a cone-beam dataset estimates dy with the other five pose parameters (--freeze dy keeps it fixed). Everything else is as for parallel scans: the coupled solver, translation seeding and saved alignments (a sixth column of the aligned file's poses, applied by tomojax recon, and dy_world in exported parameter files).

Two gauges apply. As in parallel beams, moving the whole volume rigidly and every pose with it predicts the same data. In addition a common dy for all views rescales the image exactly as a larger object would, so the mean dy is always fixed at zero.

A centre-of-rotation offset in a cone beam is a lateral offset of the rotation axis, not a detector shift: ConeBeam.axis_offset places the axis at x = axis_offset. On cone data, --mode cor calibrates it together with the detector roll, and cor-then-pose and full calibrate both before their pose stages; the aligned file records the calibrated beam, so tomojax recon on it uses them. The calibration (calibrate_cone_axis) reconstructs thin FDK slabs near the centre, top and bottom of the volume for trial offsets, coarse to fine on binned data, and keeps the sharpest. A rolled detector makes the sharpest offset change linearly with height, which gives the roll. Axis direction stages are skipped for cone data. In Python:

import tomojax as tj

result = tj.align(scan, mode="cor")      # or "cor-then-pose" to align the poses too
recon = tj.reconstruct(result.scan)      # result.scan carries the calibrated beam

tomojax.recon.calibrate_cone_axis runs the calibration alone on a geometry and arrays. Of the six pose parameters, phi about the rotation axis is the weakest; the measurements give the accuracy of both the calibration and the pose alignment.

Multi-orbit ConeSegments scans and scans that already carry poses (an ASTRA import, say, or an earlier alignment) are aligned in pose mode only, as one scan on top of their poses; see multi-orbit scans.

Where the aligned volume sits

The projections cannot tell where the object is: rotating or shifting the whole object, and every view's pose by the opposite motion, predicts exactly the same data. Left alone, the solver can end anywhere along that motion, and the volume comes out displaced.

tj.align, align_multires and tomojax align therefore report the estimate with the least per-view motion: the poses keep no common rotation and no rigid shift, the volume is moved to match, and info["gauge"] records what was removed. In a parallel beam with detector-frame poses, a schedule that reports the detector centre also moves the poses' constant u shift into it. To compare with a known truth, take its least-motion version too: least_motion_estimate in tomojax.alignment.api moves any volume and pose table to it.

The motion is a symmetry only while the object stays inside the grid. If moving the volume would push more than 1% of it out of the grid, the grid edge already pins the estimate, which is then returned unchanged.

Choose the translation frame

tj.align and tomojax align use detector-frame translations (pose_translation_frame = "detector"). A bare AlignConfig() defaults to pose_translation_frame="object": T_nominal @ se3_from_pose_params(params), with translations as physical lengths along the object's x and z axes. Near a 90-degree view those two directions project onto nearly the same detector direction, so this representation cannot express every image-plane displacement. A constant detector shift, such as a centre-of-rotation offset, then needs enormous translations near those views. On the free-voxel pilot, the chip phantom and the analytic 128³ scans, detector-frame recovery was as accurate as object-frame recovery or better (measurements). The setting pose_translation_frame = "object" in a --config file restores the older tables. The aligned file records the frame, and tomojax recon applies the poses in it.

In the detector frame, rotations still compose in the object frame. dx and dz add physical lengths along lab x and z after the nominal transform; nominal translation along the beam is preserved. Multiply pixel displacements by the detector spacing before supplying initial parameters. If the detector grid is rolled, these remain lab x and z directions rather than rolled sensor-column and row directions.

The frame also applies to align_multires, pose smoothness models, reconstruction and setup objectives. Checkpoints carry the frame, and resuming with a different frame raises an error.

tj.load("aligned.nxs").poses reads the poses tomojax align saved. To export them as JSON (save_alignment_params_csv writes CSV the same way, with a frame column for detector-frame poses), pass the same frame:

from tomojax.alignment.api import save_alignment_params_json

result = tj.align(scan, mode="pose")
save_alignment_params_json(
    "poses.json", result.poses, du=scan.detector.du, dv=scan.detector.dv,
    translation_frame="detector",
)

Expert settings for hard cases

ray_integrator = "exact" in a --config file (in Python, AlignConfig(ray_integrator="exact")) integrates the zero-extended trilinear voxel interpolant exactly between voxel-centre planes, using two-point Gaussian quadrature. It is 10–30 times slower than the default Joseph sampling. It supports rigid parallel-ray poses, including tilted scans, anisotropic voxels, shifted detector centres and irregular view angles. Reconstruction, pose loss, setup validation and translation seeds use the same operator. The CLI and coupled_pose_config default to "joseph"; AlignConfig() itself keeps "sampled". A checkpoint cannot resume with a different choice. Exact polynomial quadrature does not model detector pixel area, beam spectrum or scatter, and does not guarantee joint recovery. On data made with the same voxel basis, it recovers poses to numerical precision; that is an inverse crime, not evidence about real data.

With exact integration, the Huber-FISTA alignment path can use CUDA forward and matched adjoint kernels with changing poses. Pose derivatives and acceptance scoring use the exact JAX reference on the selected JAX device. The raw exact CUDA calls do not provide automatic differentiation; differentiable reconstruction layers use the JAX reference.

The default Jacobian differentiates within the current interpolation cells, which sharp voxel boundaries can make unsuitable for a pose update. The alignment solver reference describes the symmetric numerical columns and coupled-solve settings, and what they cost.

For calibrated laminography scans, region = tj.estimate_region(aligned.scan) selects an axial region from coarse projection-validation probes; reconstruct region.scan and inspect region.info for uncertainty and any memory-driven coarsening. Ambiguous or multilayer data retain a broad survey, not an assumed thin slab; the region does not certify physical support or reconstruction accuracy.